Specifications
Criterion® 60
Product Specifications
Dimensions: 188mm x 198mm x 94mm
Weight: 680g
Tubing Fittings: Female Luer
Display Screen: 4x20 LCD Display
Performance Specifications
Pressure Bar Graph: Range: -10 to 39.5cm H2O Resolution:
0.5cm H2O
High Pressure Alarm: Range: 5 to 99cm H2O, Off Resolution:
1.0cm H2O
Low Pressure Alarm: Range: Off, 1 to 20cm H2O Resolution:
1.0cm H2O
Peak Pressure Display: (Ppk) Range: 0 to 99.5cm H2O Resolution:
0.5cm H2O
O2concentration:(%) Range: 6 to 100%. Resolution:
0.5% O2
High O2% Alarm: Range: 19 to 100%, Off . Resolution:
1.0%.
Low O2% Alarm: Range: 18 to 99%. Resolution:
1.0%.
O2 Sensor Response Time: 90% in 7 seconds at room temperature  
O2 Sensor Life: Over 1 year under normal conditions.  
Operating Temperature 5° to 45° C at 15 to 95% relative humidity  
Storage Temperature: -40° to 60° C at 95% relative humidity  
Atmospheric Pressure: 600 to 1100 hPa  
System Accuracy, Pressure: ± (1+ 3% of reading) rounded up to the nearest 0.5cm H2O  
System Accuracy, Oxygen: ± 1% when calibrated to ambient pressure  

Top Of Page